[BREAK Wakeup A]
WorkHorse Broadband ADCP Version 16.28
RD Instruments (c) 1996-2005
All Rights Reserved.
ERR:    Invalid checksum in Active Fluxgate Calibration Matrices.
        Rcvd=00002B4B, Calc=00002ABC
ERR:    Attempting to use Factory Fluxgate Calibration.
>ts?
TS = 18/08/13,05:05:47 --- Time Set (yr/mon/day,hour:min:sec)
>rf
RF = 11756888,243806208 --- Rec space used (bytes), free (bytes)

>deploy?
Deployment Commands:
RE ----------------------- Recorder ErAsE
RN ----------------------- Set Deployment Name

WD = 111 100 000 --------- Data Out (Vel,Cor,Amp; PG,St,P0; P1,P2,P3)
WF = 0176 ---------------- Blank After Transmit (cm)
WN = 030 ----------------- Number of depth cells (1-128)
WP = 00045 --------------- Pings per Ensemble (0-16384)
WS = 0400 ---------------- Depth Cell Size (cm)
WV = 175 ----------------- Mode 1 Ambiguity Vel (cm/s radial)

BP = 000 ----------------- Bottom Track Pings per Ensemble

TE = 01:00:00.00 --------- Time per Ensemble (hrs:min:sec.sec/100)
TF = **/**/**,**:**:** --- Time of First Ping (yr/mon/day,hour:min:sec)
TP = 01:20.00 ------------ Time per Ping (min:sec.sec/100)
TS = 18/08/13,05:06:55 --- Time Set (yr/mon/day,hour:min:sec)

EA = +00000 -------------- Heading Alignment (1/100 deg)
EB = +00000 -------------- Heading Bias (1/100 deg)
ED = 00000 --------------- Transducer Depth (0 - 65535 dm)
ES = 35 ------------------ Salinity (0-40 pp thousand)
EX = 11111 --------------- Coord Transform (Xform: Type,Tilts,3 Bm,Map)
Press any key to continue
                       
EZ = 1111101 ------------- Sensor Source (C,D,H,P,R,S,T)

CF = 11111 --------------- Flow Ctrl (EnsCyc;PngCyc;Binry;Ser;Rec)
CK ----------------------- Keep Parameters as USER Defaults
CR # --------------------- Retrieve Parameters (0 = USER, 1 = FACTORY)
CS ----------------------- Start Deployment

>system?
System Control, Data Recovery and Testing Commands:
AC ----------------------- Output Active Fluxgate & Tilt Calibration data
AF ----------------------- Field calibrate to remove hard/soft iron error
AR ----------------------- Restore factory fluxgate calibration data
AX ----------------------- Examine compass performance
AZ ----------------------- Zero pressure reading

CB = 811 ----------------- Serial Port Control (Baud; Par; Stop)
CP # --------------------- Polled Mode (0 = NORMAL, 1 = POLLED)
CZ  ---------------------- Power Down Instrument

FC ----------------------- Clear Fault Log
FD ----------------------- Display Fault Log

OL ----------------------- Display Features List

PA ----------------------- Pre-Deployment Tests
PC1 ---------------------- Beam Continuity
PC2 ---------------------- Sensor Data
PS0 ---------------------- System Configuration
PS3 ---------------------- Transformation Matrices

RR ----------------------- Recorder Directory
Press any key to continue
                       
RF ----------------------- Recorder Space used/free (bytes)
RY ----------------------- Upload Recorder Files to Host

>ps0
  Instrument S/N:  2269
       Frequency:  307200 HZ
   Configuration:  4 BEAM, JANUS
     Match Layer:  10
      Beam Angle:  20 DEGREES
    Beam Pattern:  CONVEX 
     Orientation:  UP  
       Sensor(s):  HEADING  TILT 1  TILT 2  TEMPERATURE
Temp Sens Offset:  -0.37 degrees C

    CPU Firmware:  16.28 [0]
   Boot Code Ver:  Required:  1.13   Actual:  1.13
    DEMOD #1 Ver:  ad48, Type:  1f
    DEMOD #2 Ver:  ad48, Type:  1f
    PWRTIMG  Ver:  85d3, Type:   6

Board Serial Number Data:
   F1  00 00 02 00 48 2E  09 REC727-1000-04A
   E3  00 00 03 01 A5 2E  09 CPU727-2000-00H
   24  00 00 03 01 A0 15  09 PIO727-3000-04C
   6C  00 00 03 01 87 4F  09 DSP727-2001-04F
>ol                             FEATURES
---------------------------------------------------------------------
Feature                                                     Installed
---------------------------------------------------------------------
Bottom Track                                                    Yes
Water Profile                                                   Yes
High Resolution Water Modes                                      No
Lowered ADCP                                                     No
Wave Gauge Acquisition                                           No
Shallow Bottom Mode                                              No
High Rate Pinging                                                No

See your technical manual or contact RDI for information on how to
install additional capability in your WorkHorse.

>pa

PRE-DEPLOYMENT TESTS

CPU TESTS:
  RTC......................................PASS
  RAM......................................PASS
  ROM......................................PASS
RECORDER TESTS:
  PC Card #0...............................DETECTED
    Card Detect............................PASS
    Communication..........................PASS
    DOS Structure..........................PASS
    Sector Test (short)....................PASS
  PC Card #1...............................NOT DETECTED
DSP TESTS:
  Timing RAM...............................PASS
  Demod  RAM...............................PASS
  Demod  REG...............................PASS
  FIFOs....................................PASS
SYSTEM TESTS:
  XILINX Interrupts... IRQ3  IRQ3  IRQ3 ...PASS
  Wide Bandwidth........................***FAIL***
  Narrow Bandwidth.........................PASS
  RSSI Filter..............................PASS
  Transmit.................................PASS
SENSOR TESTS:
  H/W Operation............................PASS
>pc2

Press any key to quit sensor display ...

All Sensors are Internal Only.
Heading   Pitch    Roll   Up/Down    Attitude Temp   Ambient Temp   PRESSURE
130.55  -0.40  -2.51     Up          20.15C        18.20C         0.0 kPa
130.06   0.43  -3.31     Up          20.15C        18.18C         0.0 kPa
130.41   1.35  -2.86     Up          20.15C        18.20C         0.0 kPa
130.99   1.52  -1.75     Up          20.16C        18.20C         0.0 kPa
129.31  10.99  -0.77     Up          20.07C        18.18C         0.0 kPa
127.28  19.07  -1.71     Up          20.11C        18.19C         0.0 kPa
129.36   4.61  -2.36     Up          20.16C        18.20C         0.0 kPa
132.08 -15.30  -2.35     Up          20.10C        18.20C         0.0 kPa
134.12 -22.46  -1.87     Up          20.12C        18.19C         0.0 kPa
132.39  -3.26  -2.89     Up          20.14C        18.19C         0.0 kPa
127.17   1.90  15.69     Up          20.15C        18.20C         0.0 kPa
125.68   0.04  18.78     Up          20.14C        18.20C         0.0 kPa
130.57  -0.97  -7.20     Up          20.12C        18.21C         0.0 kPa
136.01   1.04 -28.50     Up          20.07C        18.20C         0.0 kPa
132.10   2.02  -6.79     Up          20.16C        18.20C         0.0 kPa
132.06   1.96  -1.68     Up          20.14C        18.21C         0.0 kPa
 85.34  -6.24  -9.85     Up          20.12C        18.21C         0.0 kPa
 81.37  -1.21  -2.69     Up          20.12C        18.21C         0.0 kPa
 82.30  -1.83  -3.46     Up          20.12C        18.20C         0.0 kPa
134.03   1.10   2.29     Up          20.16C        18.21C         0.0 kPa
146.38  -1.65  -4.97     Up          20.15C        18.19C         0.0 kPa
186.33   1.96   0.95     Up          20.14C        18.21C         0.0 kPa
186.32   2.42  -0.59     Up          20.16C        18.21C         0.0 kPa
185.98   2.58  -1.29     Up          20.16C        18.20C         0.0 kPa
186.55     -2.82     Up          20.15C        18.23C         0.0 kPa
242.92   1.02   1.33     Up          20.16C        18.21C         0.0 kPa
243.96   0.42   1.02     Up          20.16C        18.21C         0.0 kPa
243.90   0.54   0.70     Up          20.09C        18.20C         0.0 kPa
242.48   1.23   0.04     Up          20.16C        18.21C         0.0 kPa
241.91   1.87  -0.07     Up          20.15C        18.21C         0.0 kPa

>pc1


BEAM CONTINUITY TEST

When prompted to do so, vigorously rub the selected
beam's face.

If a beam does not PASS the test, send any character to
the ADCP to automatically select the next beam.

Collecting Statistical Data...
  40  42  39  40
  41  42  39  40
  41  41  38  40
  41  42  38  39
  41  42  38  39
  41  42  38  39
  41  42  38  39
  40  42  38  39
  40  42  38  39
  40  42  38  39
  40  42  38  39
  40  42  38  39
  40  42  38  39
  40  42  38  39
  40  42  38  39
  40  42  38  39
  40  42  38  39
  40  42  38  39
  40  42  38  39
  40  42  38  39
  40  42  38  39
  40  42  38  39
  40  42  38  39
  40  42  38  39
  40  42  38  39
  40  42  38  39
  40  42  38  39
  40  42  38  39
  40  42  38  39
  40  42  38  39
  40  42  38  39
  40  42  38  39
  40  42  38  39
  40  42  38  39
  40  42  38  39
  40  42  38  39
  40  42  38  39
  40  42  38  39
  40  42  38  39
  40  42  38  39
  40  42  38  39
  40  42  38  39
  40  42  38  39
  40  42  38  39
  40  42  38  39
  40  42  38  39
  40  42  38  39
  40  42  38  39
  40  42  38  39
  40  42  38  39
  40  42  38  39
  40  42  38  39
  40  42  38  39
  40  42  38  39
  40  42  38  39
  40  42  38  39
  40  42  38  39
  40  42  38  39
  40  42  38  39
  40  42  38  39
  40  42  38  39
  40  42  38  39
  40  42  38  39
  40  42  38  39
  40  42  38  39

Rub Beam 1 = PASS
Rub Beam 2 = PASS
Rub Beam 3 = PASS
Rub Beam 4 = PASS

>pt2

  Ambient  Temperature =    18.23 Degrees C
  Attitude Temperature =    20.22 Degrees C
  Internal Moisture    = 8DEFh
>pt3

Correlation Magnitude: Wide Bandwidth

               Lag  Bm1  Bm2  Bm3  Bm4
                 0  255  255  255  255
                 1  185  192  190  191
                 2   74   97   89   85
                 3   40   68   47   48
                 4   48   55   29   43
                 5   36   45   20   30
                 6   27   39   19   23
                 7   27   31   20   21

  High Gain RSSI:    41   42   39   38
   Low Gain RSSI:    12   10    8   13

  SIN Duty Cycle:    50   49   51   47
  COS Duty Cycle:    50   51   50   50

Receive Test Results = $00000000 ... PASS
>pt4

IXMT    =      2.8 Amps rms  [Data=f3h]
VXMT    =    151.0 Volts rms [Data=ffh]
   Z    =     54.3 Ohms
Transmit Test Results = $0 ... PASS
>pt5

    0    0    0    0
    0    0    0    0
    0    0    0    0
    0    0    0    0
  255  255  255  255
    0    0    0    0
    0    0    0    0
  255  255  255  255
    0    0    0    0
    0    0    0    0
    0    0    0    0
  255  255  255  255
Electronics Test Results = $00000000
>pt6
Receive Bandwidth:
    Sample      bw    bw    bw    bw    bw
      rate  expect   Bm1   Bm2   Bm3   Bm4
       307     120    68    59    68    63 Khz
   results          PASS  FAIL  PASS  PASS
>pt7
RSSI Time Constant:

RSSI Filter Strobe 1 =   38400 Hz
  time   Bm1   Bm2   Bm3   Bm4
  msec  cnts  cnts  cnts  cnts
     1     7     8     8     7
     2    13    15    14    14
     3    18    21    18    19
     4    22    25    23    23
     5    26    29    26    27
     6    28    31    28    29
     7    31    34    31    32
     8    33    36    32    33
     9    35    37    34    35
    10    36    38    35    36
   nom    42    43    39    41
result    PASS  PASS  PASS  PASS
>rr
Recorder Directory:
Volume serial number for device #0 is 882f-e9b6

  2269D000 000        10010624 05-07-18  11:57:46p r  a [    2]
  _RDI_000 000         1746264 08-13-18   4:19:26a r  a [ 2446]

  Bytes used on device #0 = 11756888
Total capacity   =  255565824 bytes
Total bytes used =   11756888 bytes in 2 files
Total bytes free =  243806208 bytes

>ps3
Beam Width:   3.7 degrees

Beam     Elevation     Azimuth
  1         -70.20      269.95
  2         -70.00       89.94
  3         -70.15        0.05
  4         -69.98      180.05

Beam Directional Matrix (Down):
  0.3387    0.0003    0.9409    0.2420  
 -0.3420   -0.0003    0.9397    0.2397  
 -0.0003   -0.3395    0.9406   -0.2418  
  0.0003    0.3422    0.9396   -0.2399  

Instrument Transformation Matrix (Down):    Q14:
  1.4651   -1.4731    0.0045    0.0016       24004  -24135      74      27  
 -0.0039   -0.0011   -1.4636    1.4702         -64     -18  -23979   24087  
  0.2672    0.2646    0.2669    0.2648        4378    4336    4373    4339  
  1.0429    1.0329   -1.0422   -1.0340       17087   16923  -17076  -16941  
Beam Angle Corrections Are Loaded.
>ps4
Ping Sequence:  W W W W W W W W W W W W W W W W W W W W W W W W W W W W W W W W W W W W W W W W W W W W W 
>ad
Display factory calibration data or active calibration data [f or a]?f


              FACTORY FLUXGATE CALIBRATION MATRICES in NVRAM
               Calibration date and time: 11/15/2001  08:42:55
                             S inverse
                                                            
     Bx      3.1419e-01 -3.1424e-01 -3.7225e-02 -6.3277e-03 
     By     -6.6312e-04  3.4251e-04 -4.1447e-01  2.4703e-03 
     Bz     -1.5481e-01 -1.7621e-01  9.8727e-03 -2.4510e-01 
     Err     3.5122e-01  3.6124e-01 -3.1342e-03 -4.8167e-01 
                                                            
                             Coil Offset
                                         
                            3.5079e+04   
                            3.4332e+04   
                            3.3455e+04   
                            3.3994e+04   
                                         
                             Electrical Null
                                     
                               33999 
                                          press any key to continue...


                    TILT CALIBRATION MATRICES in NVRAM
                Calibration date and time: 11/15/2001  08:38:40
              Average Temperature During Calibration was   23.2 C

                   Up                              Down

                                                                   
 Roll     -2.8758e-07  -1.8271e-05        -1.0000e-07   1.8917e-05 
 Pitch    -1.7960e-05   1.7869e-07        -1.8830e-05  -2.1112e-07 
                                                                   

                                                                   
 Offset    3.1850e+04   3.0871e+04         3.1818e+04   2.9773e+04 
                                                                   

                                    
                      Null    33415 
                                    





>ad
Display factory calibration data or active calibration data [f or a]?a
ERR:    Invalid checksum in Active Fluxgate Calibration Matrices.
        Rcvd=00002B4B, Calc=00002ABC
Cannot Read Fluxgate Data from NVRAM.
Error Reading Fluxgate Calibration Matrices.
Press any key to continue ...

                    TILT CALIBRATION MATRICES in NVRAM
                Calibration date and time: 11/15/2001  08:38:40
              Average Temperature During Calibration was   23.2 C

                   Up                              Down

                                                                   
 Roll     -2.8758e-07  -1.8271e-05        -1.0000e-07   1.8917e-05 
 Pitch    -1.7960e-05   1.7869e-07        -1.8830e-05  -2.1112e-07 
                                                                   

                                                                   
 Offset    3.1850e+04   3.0871e+04         3.1818e+04   2.9773e+04 
                                                                   

                                    
                      Null    33415 
                                    





>ax

------------------------------------------------------------------------------
                    RDI Compass Error Estimating Algorithm


   Press any key to start taking data after the instrument is setup.



Rotate the unit in a plane until all data samples are acquired...
   rotate less than 5/sec.  Press Q to quit.

 N       NE       E        SE       S       SW        W        NW       N
 ^                ^                 ^                 ^                 ^
                                                        *******************************************************************************
Accumulating data ...
                                                                  
                                                                  
Calculating compass performance ...


                          >>> Total error:   4.9 <<<

Press D for details or any other key to continue...

                                                               










HEADING ERROR ESTIMATE FOR THE CURRENT COMPASS CALIBRATION:

   OVERALL ERROR:
      Peak Double + Single Cycle Error (should be < 5):   4.90

   DETAILED ERROR SUMMARY:
      Single Cycle Error:                                  4.80
      Double Cycle Error:                                  0.46
      Largest Double plus Single Cycle Error:              5.26
      RMS of 3rd Order and Higher + Random Error:          0.71

      Orientation:     Up
      Average Pitch:  -0.01         Pitch Standard Dev:    1.57
      Average Roll:   -0.79         Roll Standard Dev:     2.04


Pitch or Roll Standard Deviation is higher than recommended (> 1.0) - 
   results may be biased.
Press any key to continue...

>ax

------------------------------------------------------------------------------
                    RDI Compass Error Estimating Algorithm


   Press any key to start taking data after the instrument is setup.



Rotate the unit in a plane until all data samples are acquired...
   rotate less than 5/sec.  Press Q to quit.

 N       NE       E        SE       S       SW        W        NW       N
 ^                ^                 ^                 ^                 ^
                    ***************************************                                                   ***************************************************
Accumulating data ...
                                                                  
                                                                  
Calculating compass performance ...


                          >>> Total error:   5.0 <<<

Press D for details or any other key to continue...

                                                               










HEADING ERROR ESTIMATE FOR THE CURRENT COMPASS CALIBRATION:

   OVERALL ERROR:
      Peak Double + Single Cycle Error (should be < 5):   5.04

   DETAILED ERROR SUMMARY:
      Single Cycle Error:                                  4.91
      Double Cycle Error:                                  0.57
      Largest Double plus Single Cycle Error:              5.47
      RMS of 3rd Order and Higher + Random Error:          0.78

      Orientation:     Up
      Average Pitch:   1.24         Pitch Standard Dev:    1.55
      Average Roll:    2.52         Roll Standard Dev:     1.71


Pitch or Roll Standard Deviation is higher than recommended (> 1.0) - 
   results may be biased.
Press any key to continue...

>cz

Powering Down