 SeaTech Fluorometer Settings

S/N 117S, 2 Oct 2001 - x
Time Constant = 3 sec. 2nd from top, out of 4 sets, is jumpered
Gain = 330, none out of 3 sets is jumpered
Full Scale Concentration = 3 ?g/l
(fxpscal43c-f, CTD002-010)
. .
.-.
. .		= 3 sec time constant, J1
. .

C	B	A
.	.	.
			=gain 330, 3 micrograms/liter
.	.	.
--------------------------------------------------------------------------------------------
S/N 117S, 18 Jun 2001  2 Oct 2001
Time Constant = 3 sec. 2nd from top, out of 4 sets, is jumpered
Gain = 33, A out of 3 sets is jumpered
Full Scale Concentration = 30 ?g/l
(fxpscal43b)
. .
.-.
. .		= 3 sec time constant, J1
. .

C	B	A
.	.	.
		I	=gain 33, 30 microg/liter
.	.	.
--------------------------------------------------------------------------------------------








SeaTech Fluorometer Settings, continued

S/N 30S, 2 Oct 2001-x, (1987 SeaTech Rev2).
Time Constant = 3 sec. 2nd from top, out of 4 sets, is jumpered
Gain = med, 3x. 2nd out of 3 sets is jumpered
Full Scale Concentration = 10 mg/m3
(for test & storage, not deployed)
30S fluor:
. .
.-.
. .		= 3 sec time constant, J1
. .

. .
.-.		= med gain, 3x
. .
--------------------------------------------------------------------------------------------
S/N 30S, 2  2 Oct 2001 am - pm, (1987 SeaTech Rev2).
Time Constant = 3 sec. 2nd from top, out of 4 sets, is jumpered
Gain = high, 10x. Top out of 3 sets is jumpered
Full Scale Concentration = 3 mg/m3
(fxpscal43a, CTD001.*, fluorometer trace is suspect)
30S fluor:
. .
.-.
. .		= 3 sec time constant, J1
. .

.-.
. .		= high gain, 10x
. .
--------------------------------------------------------------------------------------------








SeaTech Fluorometer Settings, continued

S/N 30S, 1 Oct 2000-2 Oct 2001, (1987 SeaTech Rev2).
(junk1.* fxps test)
Time Constant = 3 sec. 2nd from top, out of 4 sets, is jumpered
Gain = low, 1x. 3rd from top, out of 3 sets is jumpered
Full Scale Concentration = 30 mg/m3

30S fluor:
. .
.-.
. .		= 3 sec time constant, J1
. .

. .
. .		= low gain, 1x
.-.

--------------------------------------------------------------------------------------------
S/N 30S, 28 Apr 2000-1 Oct 2001, (1987 SeaTech Rev2).
6 July 2001 flash lamp replaced, full calibration
18 Jun 2001 sent to WetLabs  repairs

Time Constant = 3 sec. 2nd from top, out of 4 sets, is jumpered
Gain = med, 3x. 2nd from top, out of 3 sets is jumpered
Full Scale Concentration = 10 mg/m3

30S fluor:
. .
.-.
. .		= 3 sec time constant, J1
. .

. .
.-.		= med gain, 3x
. .

--------------------------------------------------------------------------------------------



SeaTech Fluorometer Settings, continued

S/N 117S, 15 Feb 2000  18 Jun 2001
Time Constant = 3 sec
Gain = 100
Full Scale Concentration = 10 ?g/l

. .
.-.
. .		= 3 sec time constant, J1
. .


C	B	A
.	.	.
I			=gain 100, 10 microg/liter
.	.	.

--------------------------------------------------------------------------------------------
S/N 28, 15 Feb 2000
Time Constant = 3 sec
Gain = med, 3x
Full Scale Concentration = 10 mg/m3

30S & 28 fluors:
. .
.-.
. .		= 3 sec time constant, J1
. .

. .
.-.		= med gain, 3x
. .


	E:\endata\EN361\CTD\Fluorset7.doc		27-06-05, 15:21